- 01
As far as we know, no standardisation committee has yet dealt with this topic. When using holders made of silicone or other steam-resistant plastics, however, care must be taken to ensure that not too much surface area of the medical devices is covered. When using nubbed mats in sterilisation containers, moisture residue under the liner must be avoided.
- 02
According to our experience, loaded mesh trays (note: recommended max. weight 10 kg) usually cool down within about 2 hours, by which time the contents are ready for use. With regard to condensation, it is also important that the container is not directly exposed to the cold after sterilisation as rapid cooling leads to an increased amount of condensate. Conversely, a dry and temperature-controlled storage room has a positive effect on preventing condensation inside the container.
- 03
- 04